Waterproof pH, Temperature, Conductivity & TDS meter
The Hanna HI991300 and HI991301 are some of the most popular pH meters on the market for environmental field use. They are waterproof and durable for use in remote areas. The display shows temperature and either pH, Conductivity (EC) or Total Dissolved Solids (TDS) simultaneously. Easily and accurately measure environmental groundwater parameters. For the rainy season
Now many states require that construction sites be tested for the character of runoff in pH and Turbidity. These tests should be taken when the rain event is generally greater than half inch(California). For actual requirements see local state mandates and USEPA Federal standard 180.1 and 40 CFR 450 A&B
 HANNA INSTRUMENTS HI991300 Manual pdf HI991301 Manual pdf
Ready Function
- To ensure your subsequent measurements
are accurate, these meters will only allow calibration when the
electrode has stabilized. - Rugged, Waterproof Construction
- The ABS housing provides the instrument
with optimum protection from harmful environments.
- Dual Level Display with
Graphic Symbols - A dual level LCD displays
pH, Conductivity or TDS and Temperature at the same time. Graphic
symbols guide the user throughout the calibration process.
- Comes with a one
Year Warranty on the meter and six month warranty on the probe.
- Waterproof pH Meter
- HI991301 provides a wider range of EC and TDS
- See Specifications (Click)
- These pH meters should be calibrated just prior to use in the field.

| SPECIFICATIONS |
HI991300 |
HI991301 |
| Range |
pH
EC
TDS
C/F
|
0.00 to 14.00 pH
0 to 3999 uS/cm
0 to 2000 ppm
0-60C (32-140F)
|
0.00-14.00 pH
0.00-19.99 mS/cm
0.00-10.00ppt
0.0-60C (32-140F)
|
| Resolution |
pH
EC
TDS
C/F
|
0.01 pH
1 microS/cm
1 ppm
0.1 C/F
|
0.01 pH
1 mS/cm
1 ppt
0.1 C/F
|
| Accuracy @20deg C |
pH
EC/TDS
C/F
|
+/- 0.01pH
2%Full Scale
+/-0.5C(+/-1.0F)
|
+/- 0.01pH
2%Full Scale
+/-0.5C(+/-1.0F)
|
| Typical EMC Deviation |
pH
EC/TDS
C/F
|
+/- 0.3pH
2%Full Scale
+/-1C(+/-1.0F)
|
+/- 0.3pH
2%Full Scale
+/-1C(+/-1.0F)
|
| EC/TDS Calibration |
|
1 point Auto @1382ppm or 1413microS/cm |
1 point Auto @6.44ppt or 12880microS/cm |
| pH Calibration |
|
Automatic 1 or 2 points with 2
sets of memorized buffers
(4.01/7.01/10.01 or 4.01/6.86/9.18) |
| EC/TDS Ratios |
|
Selectable between 0.45, 0.50 (default),
0.55, 0.60, 0.65, 0.68, 0.70, 0.75, 1.00 PPM=1 µS |
Temperature
Compensation |
|
pH: Automatic from 0 to 60°C;
EC/TDS: Automatic from 0 to 60°C with selectable B
of 0.0, 1.8, 1.9 (default), 2.0, 2.1, 2.2, 2.3, 2.4% per °C |
| Probe |
|
HI 1288 pH/EC/TDS
with temp sensor, DIN connector and 1 m (3.3) cable (included) |
Battery Life/Type
Environment
Dimensions
Weight |
|
500 hours of continuous use/4 x
1.5V AAA
0 to 50°C (32 to 122 °F); RH 100%
143 x 80 x 38 mm (5.6 x 3.1 x 1.5)
320g (11.3 oz) |
HI 991300(1) is supplied with pH/EC/TDS/°C probe,
batteries and instructions in a hard carrying case.
|