| Size |
13" (33 cm) long,
10.5" (26.5 cm) wide, 5" (12.5 cm) high |
| Weight: |
14.6 lb. (6.6 kg) |
| Keypad |
5 dedicated keys (Power, Start/Stop, Calibrate, Enter and Exit) and three menu keys |
| Display |
Six Line Graphical
LCD |
| Battery Capacity |
<6 hours/ 7.5 hours (8 hours with mains warm up) |
| Serial Output |
RS-232, 38.4K baud, 8 data bits, no parity for connection
to Windows based PC |
| Detector |
Photoionization detector with standard
10.6 eV UV lamp |
| Audio Output |
95 decibels @ 2048 Hz, on Alarm |
| Operating
Temp. Range |
0°C to 40°C (32°F to 105°F) |
| Benzene Analysis Time |
<60 seconds |
| BTEX Analysis Time |
<250 seconds |
| Detection Limits |
Benzene 50PPB LDL / 100PPM UDL
Toluene 50PPB LDL / 200PPM UDL
Ethylbenzene 100PPB LDL / 200PPM UDL
m-Xylene 100PPB LDL / 200PPM UDL |
| Power |
10-18 VDC, 115 or 240
VAC, adapter provided |
| Intrinsic Safety |
Class I, Division 1,
Groups A, B, C, & D pending
ATEX II 2 G EEx ib IIC T4 (TA = 0°C to +40°C) pending |
| Operating Humidity Range |
0 to 95% relative humidity (non-condensing) |
| Carrier Gas |
Built-in cylinder for Ultra Zero Air, 8 hours
continuous operation |
Software
Site Chart |
Photovac
Site Chart Software uses a PC or laptop for data storage, data
manipulation, and Application Assay customization. With Site Chart
software, all assays are easily customized for specific applications.
Additional compounds can be added to the assay. Customized,
site-specific application assays are readily saved, reloaded,
and reused. |