| Sensor Specifications |
| Gas Monitor |
Range |
Resolution |
Response /
Time T90 |
| VOCs |
0 to 9999
10 to 999 ppm
1000 to 9999 ppm |
ppb 1 ppb
0.1 ppm
1 ppm
|
< 3 s < 3 s < 3 s |
| Detector Specifications |
| Size |
10" L x 3.0" W x 2.5" H (25.5 cm x 7.6 cm x 6.4 cm) |
| Weight |
26 oz (738 g) |
| Sensors |
Photoionization sensor standard 10.6 eV or optional 9.8 eV or 11.7 eV lamps |
| Battery |
Rechargeable, external field-replaceable Lithium-Ion battery pack |
| Operating Hours |
16 hours of operation (12 hours with alkaline battery) |
| Display |
Graphic 4 lines, 28 x 43 mm, w/LED backlight for enhanced display readability |
| Keypad |
1 operation and 2 programming keys, 1 flashlight on/off |
| Direct Readout Instantaneous reading |
• VOCs as ppm by volume
• High values
• STEL and TWA
• Battery and shutdown voltage
• Date, time, temperature |
| Alarms |
95 dB (at 30cm) buzzer, flashing red LED to indicate
exceeded preset limits • STEL and TWA: 1 beep and flash per second
• Alarms latching with manual override or automatic reset
• Additional diagnostic alarm and display message for low |
| EMI/RFI |
Highly resistant to EMI/RFI |
| Compliant with EMC Directive 89/336/EEC |
| IP Rating |
• IP67 unit off and without flexible probe
• IP65 unit running |
| Datalogging |
Standard 6 months at one-minute intervals |
| Calibration |
Two-point or three-point calibration for zero and span Calibration memory for 8 calibration gases, alarm limits,
span values and calibration dates |
| Sampling Pump |
• Internal, integrated flow rate at 400 cc/mn
• Sample from 100' (30m) horizontally and vertically
Low Flow Alarm • Auto pump shutoff at low-flow condition |
| Communication |
• Download data and upload instrument set-up
from PC through charging cradle or optional Bluetooth™
• Wireless data transmission through built-in RF modem
Frequency 902 to 928 MHz (license-free), 2.400 to 2.4835 GHz
(license-free), 433 MHz, 869 MHz
RF Range Up to 500' (900 MHz, 433 Mhz, 869 Mhz),
extendable with RAELink3 Repeater to 2 miles
Hazard Area • US and Canada: UL, cUL, Classified as Intrinsically Safe
Approval for use in Class I, Division I Groups A, B, C, D
• Europe: ATEX II 1G EEx ia IIC T4 (pending)
• IECEx: II 1G EEx ia IIC T4 (pending) |
| Temperature |
-4° to 113° F (-20° to 50° C) |
| Humidity |
0% to 95% relative humidity (non-condensing) |
| Warranty |
Lifetime on non-consumable components (per RAE Systems
Standard Warranty), 3-year warranty for 10.6 eV lamp,
1 year for pump and battery |
| *Specifications are subject to change |